High-performance Optical Emission Spectrometer (OES) with top-tier accuracy, precision, and low detection limits for 60+ elements, including O, N and H analysis.
Advanced Optical Emission Spectrometer for accurate elemental analysis down to low and single-ppm detection limits.
Highlights of Metavision-1008i3
-
Analyse 60+ elements in standard calibration, including low N, O, C, H, etc.
-
Low & single PPM detection limits.
-
Hermetically sealed, cooled dual optics.
-
Multi-stage internal argon repurification for the DUV optics.
-
Current controlled digital PWM source for plasma generation.
-
Soluble-insoluble analysis (Al, Ti, B, Ca, etc.).
-
Single-sample re-standardisation/re-calibration.
Features of Metavision-1008i3
-
State-of-the-art Optics: The Metavision-1008i3 features ultra-high-resolution dual optics with the latest generation CMOS detectors and custom-developed optical components to ensure the highest levels of accuracy and stability. With a wavelength range of 118-671 nm (extendable to 800 nm) and featuring an RLD of 0.8 nm/mm, it also offers class-leading optical resolution, which directly translates into lower detection limits, higher stability, and better elemental coverage than any OES in its class.
-
High Purity Metal Analysis: The Metavision-1008i3 ensures low-level detection of key trace and alloying elements, including C, S, P, B, Ni, Cr, Li, and Na, as well as 99.995%+ purity analysis.
-
Exceptional Gaseous Element Analysis: Accurately detects Nitrogen in all forms of Steels down to 8 ppm, Oxygen in Steel, Nickel and Copper down to 10 ppm, Oxygen and Hydrogen in Titanium down to 10 ppm and 5 ppm, respectively.
-
Soluble-Insoluble Analysis: The Metavision-1008i3 achieves accurate analysis of acid-soluble and acid-insoluble components of Al, Ti, B and Ca in Steels.
-
SmartSTD: Single Sample Re-standardisation Advantage: The Metavision-1008i3 reduces re-standardisation time to under 5 minutes using a single sample, increasing productivity by up to 85% and reducing both operational costs and efforts.
Special Applications of Metavision-1008i3
The Metavision-1008i3 enables users to resolve the widest range of applications across various industries, ensuring high accuracy, reliability, and repeatability.
-
Wire rods: 3 mm to 13 mm diameter.
-
Mini samples: 4 mm to 13 mm sample fit diameter.
-
Master alloy analysis: Ultra-high concentration of key alloying elements.
-
Ultra-fine wires: 0,1 mm to 3 mm diameter.
-
Thin foils: 0,1 mm to 1 mm thickness.
Ecosystem
A comprehensive ecosystem for your Optical Emission Spectrometer.
-
Wireless RTDS: Wirelessly transmit your readings from the lab to the melting platform.
-
ArmourSafe: Protect your OES against unstable power and temperature for optimal performance.
-
MetaLib Pro: The world's largest library for instant metal-grade identification.
-
MPALabTab: Mobile application to access your OES data from anywhere, anytime.
-
LIMS: Advanced data integration and analysis tool.
-
Charge Correction: Enhance your melt recipe in real-time with melt process optimisation tool.
-
Sample Preparation Machine: Sample preparation machine for grinding, polishing, turning, and electrode pointing in ferrous and non-ferrous applications.
-
CleanSharp: Brushless electrode cleaner-sharpener to make the electrode tip cleaning and pointing easy, safe, and swift.
Tech Specifications of Metavision-1008i3
General and Optics:
-
Technology: Arc/Spark Optical Emission
-
Form Factor Options: Benchtop/Floor-standing
-
Configuration: Dual vacuum-free optics
-
Design: Rowland Circle with Paschen-Runge mountings
-
Detector Type: CMOS/CCD
-
Detector Arrangement: Multi-Detector Array
-
Wavelength Coverage: 118-671 nm (optionally to 800 nm)
-
Focal Length: 400 mm
-
Grating (grooves/mm): 3,600 & 2,400
-
Reciprocal Linear Dispersion: 0.8 nm/mm
-
Internal Argon Purification: Multi-stage purification
-
Optics features: Peltier Cooled Optics, Automated Profiling, Thermal Stabilization
Excitation Source:
-
Type: Fully Digital PWM.
-
Control Metric: Current-controlled.
-
Max. Discharge Frequency: 1.000 Hz.
-
Input Voltage range: 90-270 VAC
-
Frequency: 50/60 Hz
Analysis Features & Applications:
-
Elemental Coverage: 60+
-
Available Factory Calibrations: Ferrous, Aluminium, Copper, Nickel, Lead, Tin, Zinc, Titanium, Magnesium, Cobalt, Zirconium and Silver.
-
Purity Analysis: 99.995%+
-
Analysis features: Hydrogen (H) in Ti; Oxygen (O) in Fe, Cu, Ti & Ni; Nitrogen (N) in Steels; Soluble-Insoluble in Steels (Al, Ti, B, Ca); Sodium (Na) & Lithium (Li) in Aluminium and RoHS Elements (Pb, Hg, Cd).
-
Standardisation/Re-calibration: Smart STD (Single-sample Technique).
-
Special Applications: Ultra-Fine Wire Analysis, Thin Foil Analysis, Thin Sheet Analysis, Side-On Wire Analysis, Wire Rod Analysis and Small-Sized Sample Analysis.
Physical Dimensions and Weights:
-
Size (Benchtop): 920mm (L) x 760mm (W) x 685mm (H)
-
Weight (Benchtop): ~98 Kgs
-
Size (Floor-Standing): 920mm (L) x 760mm (W) x 1360mm (H)
-
Weight (Floor-Standing): ~210 Kgs
Environmental Conditions:
-
Operating Temperature: 18 - 35˚C
-
Storage Temperature: -10 - 50˚C
-
Relative Humidity: 20 - 80% (Non-condensing)
Certifications and Compliance:
-
Regulatory Certifications: CE Certified.
- Test Method References: ASTM, DIN.